EasyManua.ls Logo

Endress+Hauser Proline Promass F 300 - Configuration Management

Endress+Hauser Proline Promass F 300
264 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning Proline Promass F 300 EtherNet/IP
160
Parameter Prerequisite Description Selection / User
entry / User
interface
Factory setting
SSID name The User-defined option is
selected in the Assign SSID
name parameter.
The WLAN access point
option is selected in the
WLAN mode parameter.
Enter the user-defined SSID
name (max. 32 characters).
The user-defined SSID
name may only be
assigned once. If the SSID
name is assigned more
than once, the devices
can interfere with one
another.
Max. 32-digit
character string
comprising numbers,
letters and special
characters
EH_device
designation_last 7
digits of the serial
number (e.g.
EH_Promass_300_A
802000)
Connection state Displays the connection status. Connected
Not connected
Not connected
Received signal strength Shows the received signal
strength.
Low
Medium
High
High
10.6.7 Configuration management
After commissioning, you can save the current device configurationor restore the previous
device configuration.
You can do so using the Configuration management parameter and the related options
found in the Configuration backup submenu.
Navigation
"Setup" menu → Advanced setup → Configuration backup
Configuration backup
Operating time
→  160
Last backup
→  160
Configuration management
→  161
Backup state
→  161
Comparison result
→  161
Parameter overview with brief description
Parameter Description User interface / Selection Factory setting
Operating time Indicates how long the device has been in
operation.
Days (d), hours (h), minutes
(m) and seconds (s)
Last backup Shows when the last data backup was saved
to HistoROM backup.
Days (d), hours (h), minutes
(m) and seconds (s)