Proline Promass F 300 EtherNet/IP Commissioning
125
Parameter Prerequisite Description Selection / User
interface / User
entry
Factory setting
Switch-on value In the Operating mode
parameter, the Switch
option is selected.
In the Switch output
function parameter, the
Limit option is selected.
Enter measured value for the
switch-on point.
Signed floating-point
number
Country-specific:
0 kg/h
0 lb/min
Switch-off value In the Operating mode
parameter, the Switch
option is selected.
In the Switch output
function parameter, the
Limit option is selected.
Enter measured value for the
switch-off point.
Signed floating-point
number
Country-specific:
0 kg/h
0 lb/min
Switch-on delay The Switch option is
selected in the Operating
mode parameter.
The Limit option is selected
in the Switch output
function parameter.
Define delay for the switch-on
of status output.
0.0 to 100.0 s 0.0 s
Switch-off delay The Switch option is
selected in the Operating
mode parameter.
The Limit option is selected
in the Switch output
function parameter.
Define delay for the switch-off
of status output.
0.0 to 100.0 s 0.0 s
Failure mode Define output behavior in
alarm condition.
Actual status
Open
Closed
Open
Invert output signal Invert the output signal. No
Yes
No
* Visibility depends on order options or device settings
10.5.10 Configuring the relay output
The Relay output wizard guides the user systematically through all the parameters that
have to be set for configuring the relay output.
Navigation
"Setup" menu → Relay output 1 to n
RelaisOutput 1 to n
Switch output function
→  126
Assign flow direction check
→  126
Assign limit
→  127
Assign diagnostic behavior
→  127