Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
201
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
383 Memory content 1. Restart device
2. Delete T-DAT via 'Reset device'
parameter
3. Replace T-DAT
0x100016E
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
387 HistoROM backup failed Contact service organization 0x1000288
Status signal F
Diagnostic behavior Alarm
12.8.3 Diagnostic of configuration
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
303 I/O 1 to n configuration changed 1. Apply I/O module configuration
(parameter 'Apply I/O
configuration')
2. Afterwards reload device
description and check wiring
0x400026C
Status signal M
Diagnostic behavior Warning
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
330 Flash file invalid 1. Update firmware of device
2. Restart device
0x40002C9
Status signal M
Diagnostic behavior Warning
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
331 Firmware update failed 1. Update firmware of device
2. Restart device
0x10002CA
Status signal F
Diagnostic behavior Warning
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
410 Data transfer 1. Check connection
2. Retry data transfer
0x100008B
Status signal F
Diagnostic behavior Alarm