Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
202
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
412 Processing download Download active, please wait 0x2000204
Status signal C
Diagnostic behavior Warning
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
431 Trim 1 to n Carry out trim 0x2000004
Status signal C
Diagnostic behavior Warning
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
437 Configuration incompatible 1. Restart device
2. Contact service
0x1000060
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
438 Dataset 1. Check data set file
2. Check device configuration
3. Up- and download new
configuration
0x400006A
Status signal M
Diagnostic behavior Warning
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
441 Current output 1 to n 1. Check process
2. Check current output settings
• 0x8000099
• 0x80000B6
Status signal S
Diagnostic behavior Warning
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
442 Frequency output 1 to n 1. Check process
2. Check frequency output settings
• 0x800008A
• 0x8000122
Status signal S
Diagnostic behavior Warning