Commissioning Proline Promass F 300 EtherNet/IP
114
10.5.9 Configuring the pulse/frequency/switch output
The Pulse/frequency/switch output wizard guides you systematically through all the
parameters that can be set for configuring the selected output type.
Navigation
"Setup" menu → Advanced setup → Pulse/frequency/switch output
‣
Pulse/frequency/switch output
1 to n
Operating mode
→ 114
Parameter overview with brief description
Parameter Description Selection Factory setting
Operating mode Define the output as a pulse, frequency or
switch output.
• Pulse
• Frequency
• Switch
Pulse
Configuring the pulse output
Navigation
"Setup" menu → Pulse/frequency/switch output
‣
Pulse/frequency/switch output
1 to n
Operating mode
→ 115
Terminal number
→ 115
Signal mode
→ 115
Assign pulse output
→ 115
Value per pulse
→ 115
Pulse width
→ 116
Failure mode
→ 116
Invert output signal
→ 116