Commissioning Proline Promass F 300 EtherNet/IP
116
Parameter Prerequisite Description Selection / User
interface / User
entry
Factory setting
Pulse width In the Operating mode
parameter, the Pulse option is
selected, and one of the
following options is selected in
the Assign pulse output
parameter (→ 115):
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow
*
• Carrier mass flow
*
Define time width of the
output pulse.
0.05 to 2 000 ms 100 ms
Failure mode In the Operating mode
parameter, the Pulse option is
selected, and one of the
following options is selected in
the Assign pulse output
parameter (→ 115):
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow
*
• Carrier mass flow
*
Define output behavior in
alarm condition.
• Actual value
• No pulses
No pulses
Invert output signal – Invert the output signal. • No
• Yes
No
* Visibility depends on order options or device settings
Configuring the frequency output
Navigation
"Setup" menu → Pulse/frequency/switch output
‣
Pulse/frequency/switch output
1 to n
Operating mode
→ 117
Terminal number
→ 117
Signal mode
→ 117
Assign frequency output
→ 118
Minimum frequency value
→ 119
Maximum frequency value
→ 119
Measuring value at minimum
frequency
→ 120