EasyManua.ls Logo

Endress+Hauser Proline Promass F 300 - Configuring the Status Input; Configuring the Current Output

Endress+Hauser Proline Promass F 300
264 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promass F 300 EtherNet/IP Commissioning
109
10.5.7 Configuring the status input
The Status input submenu guides the user systematically through all the parameters that
have to be set for configuring the status input.
Navigation
"Setup" menu → Status input
Status input 1 to n
Assign status input
→  109
Terminal number
→  109
Active level
→  109
Terminal number
→  109
Response time status input
→  109
Terminal number
→  109
Parameter overview with brief description
Parameter Description User interface / Selection / 
User entry
Factory setting
Terminal number Shows the terminal numbers used by the
status input module.
Not used
24-25 (I/O 2)
22-23 (I/O 3)
Assign status input Select function for the status input. Off
Reset totalizer 1
Reset totalizer 2
Reset totalizer 3
Reset all totalizers
Flow override
Off
Active level Define input signal level at which the
assigned function is triggered.
High
Low
High
Response time status input Define the minimum amount of time the
input signal level must be present before the
selected function is triggered.
5 to 200 ms 50 ms
10.5.8 Configuring the current output
The Current output wizard guides you systematically through all the parameters that have
to be set for configuring the current output.

Table of Contents

Related product manuals