Commissioning Proline Promass F 300 EtherNet/IP
110
Navigation
"Setup" menu → Current output
‣
Current output 1 to n
Terminal number
→ 110
Signal mode
→ 110
Assign current output 1 to n
→ 111
Current span
→ 112
0/4 mA value
→ 112
20 mA value
→ 112
Fixed current
→ 112
Damping output 1 to n
→ 112
Failure mode
→ 113
Failure current
→ 113
Parameter overview with brief description
Parameter Prerequisite Description User interface /
Selection / User
entry
Factory setting
Terminal number – Shows the terminal numbers
used by the current output
module.
• Not used
• 24-25 (I/O 2)
• 22-23 (I/O 3)
–
Signal mode – Select the signal mode for the
current output.
• Passive
• Active
Active