Loading...
Proline Promass F 300 EtherNet/IP Technical data
231
Configuration of the device
address
DIP switches on the electronics module for IP addressing (last octet)
DHCP
Manufacturer-specific software (FieldCare)
Add-on Profile Level 3 for Rockwell Automation control systems
Web browser
EtherNet/IP tools, e.g. RSLinx (Rockwell Automation)
Device Level Ring (DLR) Yes
System integration Information regarding system integration →  80.
Cyclic data transmission
Block model
Input and output groups
16.5 Power supply
Terminal assignment →  36
Device plugs available →  36
Pin assignment, device plug →  36
Supply voltage
Order code for
"Power supply"
terminal voltage Frequency range
Option D DC24 V ±20%
Option E AC100 to 240 V –15…+10% 50/60 Hz
Option I
DC24 V ±20%
AC100 to 240 V –15…+10% 50/60 Hz
Power consumption Transmitter
Max. 10 W (active power)
Current consumption Transmitter
Max. 400 mA (24 V)
Max. 200 mA (110 V, 50/60 Hz; 230 V, 50/60 Hz)
Power supply failure Totalizers stop at the last value measured.
Configuration is retained in the plug-in memory (HistoROM DAT).
Error messages (incl. total operated hours) are stored.
Electrical connection →  37