Technical data Proline Promass F 300 EtherNet/IP
252
Radio approval The measuring device has radio approval.
For detailed information on the radio approval, see the Special Documentation
→ 256
Additional certification Marine approval
Currently valid certificates are available:
• In the Download Area of the Endress+Hauser Internet site: www.endress.com →
Downloads
• Specify the following details:
Search area: Approval & Certificates→ Marine
CRN approval
Some device versions have CRN approval. A CRN-approved process connection with a CSA
approval must be ordered for a CRN-approved device.
Tests and certificates
• Pressure test, internal procedure, inspection certificate
• EN10204-3.1 Material certificate, wetted parts and secondary containment
• PMI test (XRF), internal procedure, wetted parts, test report
• EN10204-2.1 confirmation of compliance with the order and EN10204-2.2 test report
Testing of welded connections
Option Test standard Component
ISO 23277 AL2x (PT)
ISO 10675-1 AL1 (RT, DR)
ASME
B31.3
NFS
ASME
VIII Div.1
Appx. 4+8
NORSOK
M-601
Measuring tube Process connection
CF x PT RT
KK x PT RT
KP x PT RT
KR x VT, PT VT, RT
K1 x PT DR
K2 x PT DR
K3 x PT DR
K4 x VT, PT VT, DR
PT = penetrant testing, RT = radiographic testing, VT = visual testing, DR = digital radiography
All options with test report
Other standards and
guidelines
• EN 60529
Degrees of protection provided by enclosures (IP code)
• IEC/EN 60068-2-6
Environmental influences: Test procedure - Test Fc: vibrate (sinusoidal).
• IEC/EN 60068-2-31
Environmental influences: Test procedure - Test Ec: shocks due to rough handling,
primarily for devices.
• EN 61010-1
Safety requirements for electrical equipment for measurement, control and laboratory
use - general requirements