Proline Promass F 300 EtherNet/IP System integration
89
Description (format) Byte Bit Value
28. External reference density unit (integer) 29-30 16 • 32840 (240): kg/Nm
3
• 32841 (240): kg/Nl
• 32842 (240): g/Scm
3
• 32843 (240): kg/Scm
3
• 32844 (240): lb/Sft
3
29. Not used 31-32 16 –
30. External temperature (real) 33-36 32 Data format:
Byte 1-4: External temperature
Floating-point number (IEEE754)
31. External temperature unit (integer) 37-38 16 • 1001 (32): °C
• 1002 (33): °F
• 1000 (35): K
• 1003 (34): °R
32. Not used 39-40 16 –
33. External value % S&W (real) 41-44 32 Data format:
Byte 1-4: External value, % S&W
Floating-point number (IEEE754)
34. External value, water cut % (real) 45-48 32 Data format:
Byte 1-4: External value, water cut %
Floating-point number (IEEE754)
35 Flow override monitoring 49-50 16 • 33004 (0): Off
• 33006 (1): On
36 Zero point adjustment monitoring 51-52 16 • 32823 (0): Cancel
• 33242 (0): Active
• 248 (0): Error zero point adjustment
• 33158 (1): Start
Permanently assigned configuration group
Config assembly (Assem104), 2704 byte
Description (format) Bits Byte Offset
1. None 32 4 0
2. Parameter 36 – Write protection 8 1 4
3. None 8 1 5
4. Parameter 87 System units Mass flow unit 16 2 6
5. Parameter 86 System units Mass unit 16 2 8
6. Parameter 93 System units Volume flow unit 16 2 10
7. Parameter 92 System units Volume unit 16 2 12
8. Parameter 80 System units Corrected volume flow unit 16 2 14
9. Parameter 79 System units Corrected volume unit 16 2 16
10. Parameter 81 System units Density unit 16 2 18
11. Parameter 89 System units Reference density unit 16 2 20
12. Parameter 91 System units Temperature unit 16 2 22
13. None 16 2 24
14. Parameter 88 System units Press. eng. unit 16 2 26
15. Parameter 85 System units Kinematic viscosity unit 16 2 28