9 Trouble-shooting PROline Prosonic Flow 93 PROFIBUS-DP/-PA
134 Endress+Hauser
Zeropoint adjustment
is not possible
P 0-ADJ. FAIL CH2
$ # 744
744 non specific UNCERTAIN
0x40
no limits
Cause:
Zero point adjustment of channel 2
is not possible or was cancelled.
Remedy:
Make sure that zero point
adjustment is carried out at “zero
flow” only (v = 0 m/s)

Page 113.
No communication to
amplifier
device failure BAD
0x0F
constant
Cause:
Communication error. No
communication to the amplifier.
Remedy:
1. Switch power supply off and
then on again
2. Check whether the electronics
boards are correctly inserted in
their holders
Page 138
Device status and
diagnosis message
(Control system)
Device status message
(Display)
No. Initial status
Analog Input Block/
Totalizer Block
Measured
value Quality /
Substatus /
Alarm limit
Cause/remedy