5 Operation PROline Prosonic Flow 93 PROFIBUS-DP/-PA
56 Endress+Hauser
5.2 Operation via the local display
5.2.1 Display and operating elements
The local display enables you to read all important parameters directly at the measuring
point and configure the device using the “Quick Setup” or the function matrix.
The display consists of four lines; this is where measured values and/or status variables
(direction of flow, bar graph, etc.) are displayed. You can change the assignment of
display lines to different variables to suit your needs and preferences (
see the
“Description of Device Functions” manual).
Fig. 55: Display and operating elements
Liquid crystal display (1)
The backlit, four-line liquid-crystal display shows measured values, dialog texts, error messages and
information messages. The display as it appears during standard measuring mode is known as the HOME
position (operating mode).
Optical sensors for Touch Control (2)
Plus/minus keys (3)
– HOME position
Direct access to totalizer values and actual values of inputs/outputs
– Enter numerical values, select parameters
– Select different blocks, groups or function groups within the function matrix
Press the +/– keys simultaneously to trigger the following functions:
– Exit the function matrix step by step
HOME position
– Press and hold down the +/– keys for more than 3 seconds
Return directly to HOME position
– Cancel data entry
Enter key (4)
–HOME position
Enter the function matrix
– Save the numerical values you input or settings you change