4 Wiring PROline Prosonic Flow 93 PROFIBUS-DP/-PA
46 Endress+Hauser
4.2 Connecting the sensor connecting cable
4.2.1 Connecting Prosonic Flow P / W / U / DDU 18 / DDU 19
#
Warning!
Risk of electric shock. Switch off the power supply before opening the device. Do not
install or wire the device while it is connected to the power supply. Failure to comply
with this precaution can result in irreparable damage to the electronics.
Risk of electric shock. Connect the protective earth to the ground terminal on the
housing before the power supply is applied.
Fig. 47: Connecting the measuring system
Fig. 48: Connecting the sensor connecting cable
See next page for graphic legend and installation instructions.
F06-9xxxxxxx-04-06-06-xx-002F06-9xxxxxxx-04-06-06-xx-003