PROline Prosonic Flow 93 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 147
10.1.7 Operating conditions
Installation
Installation instructions Any orientation (vertical, horizontal)
Restrictions and additional installation instructions
Page 14 ff.
Inlet and outlet run Clamp On version
Page 15
Insertion version
Page 16
Length of connecting
cable
Shielded cables are offered in the following lengths:
5 m, 10 m, 15 m and 30 m
Route the cable well clear of electrical machines and switching elements.
Environment
Ambient temperature
range
Transmitter Prosonic Flow 93:
20...+60 °C
optionally: -40...+60°C
! Note!
At ambient temperatures below -20°C the readability of the display may be impaired.
Flowrate measuring sensors Prosonic Flow P (Clamp On):
–40...+80 °C / 0...+170 °C
Flowrate measuring sensors Prosonic Flow W (Clamp On):
–20...+80 °C
Flowrate measuring sensors Prosonic Flow U (clamp on):
–20...+60 °C
Flowrate measuring sensors Prosonic Flow W (Insertion version):
–40...+80 °C
Sound velocity measuring sensors DDU 18:
–40...+80 °C / 0...+170 °C
Wall thickness measuring sensor DDU 19:
0...+60 °C
Sensor cable PTFE: –40...+170 °C; sensor cable PVC: –20...+70 °C
In heated piping or piping conveying cold fluids, it is always permissible to insulate
the piping completely with the mounted ultrasonic sensors.
Install the transmitter at a shady location. Avoid direct sunlight, particularly in warm
climatic regions.
Storage temperature The storage temperature corresponds to the ambient temperature range of the
measuring transmitter and the appropriate measuring sensors and the corresponding
sensor cable (see above).