PROline Prosonic Flow 93 PROFIBUS-DP/-PA 4 Wiring
Endress+Hauser 53
Flowrate measuring sensors U (Clamp on)
The flowrate measuring sensors U fulfill all the requirements for IP 52 degree of protec-
tion. Compliance with the following points is mandatory following installation in the field
or servicing, in order to ensure that IP 52 protection is maintained:
Only use cables supplied by E+H with the corresponding sensor connectors.
The BNC cable connectors (1) must be clean, dry and undamaged.
Insert the BNC cable connectors (1), do not cant and then tighten them to the stop.
Abb. 53: Installation instructions for IP 52 degree of protection for BNC sensor connectors
1
F06-9xxxxxxx-04-05-00-xx-001