PROline Prosonic Flow 93 PROFIBUS-DP/-PA 6 Commissioning
Endress+Hauser 83
!
Note!
The display returns to the cell SETUP SENSOR (1001) if you press the ESC key
combination during interrogation.
➀
Selection of the system units only influences the functions UNIT TEMPERATURE (0422),
UNIT LENGTH (0424) and UNIT VELOCITY (0425).
➁
If a channel is selected for which a Quick Setup has already been executed, the
previous values are overwritten.
➂
During each run, all the options can be selected. If settings were made during a
previous run, these are overwritten.
➃
Selection “YES”: The value measured during Quick Setup is accepted in the appropriate
function. Selection “NO”: The measurement is discarded and the original value remains.
➄
The function SENSOR POSITION (6884) only appears when the CLAMP ON option is
set in the MEASUREMENT function and the number of traverses is 2 or 4 in the function
SENSOR CONFIGURATION (6882).
➅
The function WIRE LENGTH (6885) only appears when the CLAMP ON option is set in
the MEASUREMENT function and the number of traverses is 1 or 3 in the function
SENSOR CONFIGURATION (6882).
➆
The function ARC LENGTH (6887) only appears when the INSERTION option is set in
the MEASUREMENT function and the DUAL PATH option is selected in the SENSOR
CONFIGURATION (6882) function.