PROline Prosonic Flow 93 PROFIBUS-DP/-PA 6 Commissioning
Endress+Hauser 83
!
Note!
The display returns to the cell SETUP SENSOR (1001) if you press the ESC key
combination during interrogation.
Selection of the system units only influences the functions UNIT TEMPERATURE (0422),
UNIT LENGTH (0424) and UNIT VELOCITY (0425).
If a channel is selected for which a Quick Setup has already been executed, the
previous values are overwritten.
During each run, all the options can be selected. If settings were made during a
previous run, these are overwritten.
Selection “YES”: The value measured during Quick Setup is accepted in the appropriate
function. Selection “NO”: The measurement is discarded and the original value remains.
The function SENSOR POSITION (6884) only appears when the CLAMP ON option is
set in the MEASUREMENT function and the number of traverses is 2 or 4 in the function
SENSOR CONFIGURATION (6882).
The function WIRE LENGTH (6885) only appears when the CLAMP ON option is set in
the MEASUREMENT function and the number of traverses is 1 or 3 in the function
SENSOR CONFIGURATION (6882).
The function ARC LENGTH (6887) only appears when the INSERTION option is set in
the MEASUREMENT function and the DUAL PATH option is selected in the SENSOR
CONFIGURATION (6882) function.