Proline Prowirl 73 Troubleshooting
Endress+Hauser 67
S
$
HART-IN: T.-OUT
# 523
The HART input is activated but Prowirl has not found a
burst telegram for some time.
!
Note!
The time limit for the occurrence of this error message is
specified in the TIMEOUT HART COM function.
• Check if the pressure, temperature or density sensor is HART-
enabled and is in the BURST mode.
• Check if the wiring has been carried out according to the
diagrams on ä 30.
S
$
SIGN DELTA HEAT
# 524
The algebraic sign of the temperature differential is
different to what Prowirl 73 expected.
• If the error message occurs when commissioning the
measurement, check the setting in the INSTALLATION POINT
function (ä 157).
• If the error message occurs during operation, check whether
the algebraic sign for the temperature differential has changed.
!
Note!
Prowirl 73 cannot take a change in the algebraic sign for
temperature measurement into account!
S
!
POS. ZERO-RET.
# 601
Positive zero return active.
"
Caution!
This message has the highest display priority.
Switch off positive zero return.
S
!
SIM. CURR. OUT
# 611
Current output simulation active. Switch off simulation.
S
!
SIM. FREQ. OUT
# 621
Simulation frequency output active. Switch off simulation.
S
!
SIM. PULSE
# 631
Pulse output simulation active. Switch off simulation.
S
!
SIM. STAT. OUT
# 641
Status output simulation active. Switch off simulation.
S
$
SIM. FAILSAFE
# 691
Simulation of failsafe mode (outputs) active. Switch off simulation.
S
!
SIM. MEASURAND
# 692
Simulation of a measured variable active (e.g. mass flow). Switch off simulation.
Type Error message / No. Cause Remedy / spare part (ä 72)