Proline Prowirl 73 Troubleshooting
Endress+Hauser 67
S
$
HART-IN: T.-OUT
# 523
The HART input is activated but Prowirl has not found a
burst telegram for some time.
!
Note!
The time limit for the occurrence of this error message is
specified in the TIMEOUT HART COM function.
Check if the pressure, temperature or density sensor is HART-
enabled and is in the BURST mode.
Check if the wiring has been carried out according to the
diagrams on ä 30.
S
$
SIGN DELTA HEAT
# 524
The algebraic sign of the temperature differential is
different to what Prowirl 73 expected.
If the error message occurs when commissioning the
measurement, check the setting in the INSTALLATION POINT
function (ä 157).
If the error message occurs during operation, check whether
the algebraic sign for the temperature differential has changed.
!
Note!
Prowirl 73 cannot take a change in the algebraic sign for
temperature measurement into account!
S
!
POS. ZERO-RET.
# 601
Positive zero return active.
"
Caution!
This message has the highest display priority.
Switch off positive zero return.
S
!
SIM. CURR. OUT
# 611
Current output simulation active. Switch off simulation.
S
!
SIM. FREQ. OUT
# 621
Simulation frequency output active. Switch off simulation.
S
!
SIM. PULSE
# 631
Pulse output simulation active. Switch off simulation.
S
!
SIM. STAT. OUT
# 641
Status output simulation active. Switch off simulation.
S
$
SIM. FAILSAFE
# 691
Simulation of failsafe mode (outputs) active. Switch off simulation.
S
!
SIM. MEASURAND
# 692
Simulation of a measured variable active (e.g. mass flow). Switch off simulation.
Type Error message / No. Cause Remedy / spare part (ä 72)