15 Maintenance, Customer Service, Warranty, Disposal
80113-101_V25087
2023-11
Chapter 15
Maintenance, Customer Service, Warranty,
Disposal
Maintenance
Modifications and repairs
Modifications and repairs must not impair the safety of the unit or cart and accessories
for the patient, user and the environment. This condition is met when changes to the
structural and functional characteristics are not detrimental to safety.
Authorized persons
Modifications and repairs may only be undertaken by Erbe or by persons expressly au-
thorized by Erbe. Erbe accepts no liability if modifications and repairs to the unit or
accessories are made by unauthorized persons. This will also invalidate the warranty.
Technical safety checks
The technical safety checks determine whether the safety and operational readiness
of the unit or the cart and accessories conform to a defined technical required status.
Technical safety checks must be performed at least once a year.
What technical safety checks must
be performed?
For this unit the following technical safety checks have been stipulated:
Checking of labels and User Manual
Visual inspection of unit and accessories for damage
Testing the grounded conductor as per IEC 62353
Leakage current testing as per IEC 62353
Functional testing of all the unit's operating and control elements
Testing the monitoring equipment
Measurement of DC resistance
Testing footswitch and fingerswitch activation
Testing the automatic start mode (for VIO 100 C only)
Measurement of the output power in the CUT and COAG operating modes
The results of the technical safety checks must be documented.
If during the technical safety checks any defects are found which might endanger pa-
tients, staff or third parties, the unit may not be operated until the defects have been
remedied by competent service technicians.
Customer service
If you are interested in a maintenance contract, please contact Erbe Elektromedizin in
Germany, or your local contact in other countries. This may be an Erbe subsidiary, an
Erbe representative or a distributor.