69 / 106
8 • Monopolar Modes
80113-101_V25087
2023-11
Chapter 8
Monopolar Modes
AUTO CUT
Properties
Reproducible, extremely tissue-sparing cuts, minimal to medium hemostasis.
Areas of use
All cutting procedures in electrically conductive tissue: e.g. muscle tissue and vascular
tissue. Dissections and cutting of fine structures.
Suitable electrodes
Needle electrodes, knife electrodes, spatula electrodes.
Technical data
HF voltage waveform unmodulated sinusoidal alternating
voltage
Rated frequency 550 kHz (at R
L
= 500 Ω) ± 10 %
Crest factor 1.6 (at R
L
= 500 Ω)
Rated load resistor 500 Ω
Max. HF peak voltage 570 Vp
HF power limitation VIO 50 C: 1 watt to 50 watts in 1 watt
steps
VIO 100 C: 1 watt to 100 watts in
1 watt steps
Max. power output at rated load resis-
tor
VIO 50 C: 50 watts ± 20 %
VIO 100 C: 100 watts ± 20 %
Max. RMS current (maximum HF output
current in normal use)
VIO 50 C: 195 mA
VIO 100 C: 224 mA