2 Safety Instructions
16 / 106
80113-101_V25087
2023-11
If the unit or cart is damaged, please contact our customer ser-
vice.
Blown power fuses may only be replaced by a competent techni-
cian.
For your safety and that of the patient: Never attempt to perform
repairs or make modifications yourself. Any modification will in-
validate liability on the part of Erbe Elektromedizin GmbH.
WARNING
Connection of unit and power supply during cleaning and disinfec-
tion
Risk of electric shock to the medical personnel!
Switch off the unit. Unplug the power plug of the unit.
WARNING
The user touches the patient and the connections on the unit at
the same time.
Patient and medical staff may be put at risk by compensating cur-
rents.
Risk of injury to patient and medical staff. The unit can fail.
Do not touch the patient and the connections on the unit at the
same time.
Fire / explosion
In electrosurgery electric sparks and arcs occur at the instrument. Flammable gases,
vapors, and liquids can be set alight or caused to explode.
WARNING
Flammable anesthetics
Risk of explosion to the patient and medical personnel! Risk of dam-
age to property.
Do not use flammable anesthetics when an operation is being
performed on the head or thorax.
If use is unavoidable, you must evacuate the anesthetics before
performing electrosurgery.
WARNING
Flammable gas mixture in TUR (Transurethral Resection) and TCR
(Transcervical Endometrial Resection)
Hydrogen and oxygen can ascend into the roof of the bladder, the up-
per part of the prostate, and the upper part of the uterus. If you resect
into this gas mixture, it could combust.
Risk of combustion to the patient!
Allow the gas mixture to escape through the resectoscope sheath.
Do not resect into the gas mixture.