12 • Error messages
94 / 106
80113-101_V25087
2023-11
System error messages
19 When the AUTO START function was assigned to the
bipolar socket, there was electrically conductive con
-
tact between the electrodes of the connected bipolar
instruments.
The error may occur, for example, when the electrodes
have direct contact or when the instrument is set on a
conductive subsurface.
Avoid conductive contact between the electrodes
when assigning the AUTO
START function.
20 to 23 A footswitch or fingerswitch was pressed during start-
up.
Do not press any footswitch or fingerswitch during
start-up.
24 to 33 A button was pressed during start-up.
The messages can also appear if the keyboard is
defective.
Do not press any button during start-up.
Switch unit off and back on. If the error persists,
inform the Erbe service department.
No. Description Action
43 System error.
The message also occurs if the unit temperature lies
outside the temperature range permissible for opera
-
tion.
If necessary acclimatize the unit according to the
specifications in the Technical Data.
Switch unit off and back on. If the error persists,
inform the Erbe service department.
60 The unit uses non-calibrated defaults for certain HF
parameters. Tissue effects may deviate from the usual.
Switch unit off and back on. If the error persists,
inform the Erbe service department.
Note: With increased attention, an ongoing operation
may continue to the end despite the error.
61 The neutral electrode monitoring system uses non-
calibrated defaults for certain HF parameters. The
sensitivity of the neutral electrode monitoring system
is increased.
Switch unit off and back on. If the error persists,
inform the Erbe service department.
Note: With increased attention, an ongoing operation
may continue to the end despite the error.
62 Customized programs and unit settings (e.g. maxi-
mum ON time) are no longer available. The unit uses
the factory settings instead.
Switch unit off and back on. If the error persists,
inform the Erbe service department.
Note: With increased attention, an ongoing operation
may continue to the end despite the error.
CAUTION! Before continuing an ongoing operation,
check the program settings provided by the unit and
adjust if necessary.
Other System error. Switch unit off and back on. If the error persists,
inform the Erbe service department.