EasyManua.ls Logo

Erbe VIO 50 - General Technical Data

Erbe VIO 50
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
95 / 106
13 General Technical Data
80113-101_V25087
2023-11
Chapter 13
General Technical Data
Power connection
Rated supply voltage 100 V to 240 V (± 10 %)
Rated supply frequency 50 / 60 Hz
Line current Max. 2.0 A
Power input in standby mode < 15 watts
Power input with max. HF output 190 watts / 200 VA
Terminal for grounding (potential equalization) Yes
Power fuses T 4 A H / 250 V
Operating mode
Intermittent operation Duty cycle 25% (10 s ON / 30 s OFF)
Output frequency
Frequency 380 kHz, RL =
Dimensions and weight
Width x height x depth 280 x 135 x 300 mm
Weight 4.0 kg
Ambient conditions for transport and storage of unit
Temperature -40 °C to + 70 °C
Relative humidity 10% – 95%
Ambient conditions for operation of unit
Temperature +10 °C to +40 °C
Relative humidity 15% – 80%, non-condensing
Air pressure 54 kPa – 106 kPa
Maximum operating altitude 5000 m above sea level