2 • Safety Instructions
18 / 106
80113-101_V25087
2023-11
WARNING
Ignition of anesthetics, skin cleansers, and disinfectants in poten-
tially explosive atmospheres
If you place the unit in a potentially explosive atmosphere, anesthet-
ics, skin cleansers, and disinfectants can ignite.
Risk of fire and explosion to the patient and medical personnel! Risk
of damage to property.
Do not place the unit in potentially explosive atmospheres.
WARNING
Application of electrosurgery using an oxygen concentration in
the tracheobronchial system that is too high
Risk of fire in the respiratory tract for the patient!
Reduce the oxygen supply below a critical threshold in accordance
with therapeutic guidelines before activating the instrument in
the tracheobronchial system.
Burns
WARNING
Damaged unit, damaged accessories, modified unit, and modified
accessories
Risk of electric shock to the patient and medical personnel!
Risk of injury to patient and medical personnel!
Risk of burns to the patient and medical personnel!
Risk of material damage.
Check the unit and accessories for damage before each use (e.g.
footswitch, instrument cables and cart).
You must not use a damaged unit or damaged accessories. Re-
place defective accessories.
If the unit or cart is damaged, please contact our customer ser-
vice.
Blown power fuses may only be replaced by a competent techni-
cian.
For your safety and that of the patient: Never attempt to perform
repairs or make modifications yourself. Any modification will in-
validate liability on the part of Erbe Elektromedizin GmbH.
WARNING
Using unsuitable accessories
Risk of injury to patient and medical personnel.
Observe the notes in the Accessories chapter.