23 / 106
2 Safety Instructions
80113-101_V25087
2023-11
WARNING
A hand-held metal instrument is touched with the active instru-
ment (electrode)
Risk of hand burns!
Do not touch a hand-held metal instrument with the active instru-
ment.
WARNING
The instrument is inserted into an activated socket. The instru-
ment is inserted into an adapter or a cable that is connected to an
activated socket.
Risk of burns to the patient and medical personnel!
Do not insert instruments into an activated socket.
Do not insert instruments into an adapter that is plugged into an
activated socket.
Do not insert instruments into a cable that is plugged into an ac-
tivated socket.
CAUTION
Activation of the unit during cleaning of an instrument
Risk of burns for the medical personnel!
Do not activate the unit while you are cleaning an instrument.
Risk of infection
WARNING
The unit is contaminated
Risk of infection to the patient and medical personnel.
Follow the instructions for cleaning and disinfecting the unit.
WARNING
Alternate use of disinfectant solutions based on different active
ingredients
The agents may affect one another. The disinfection effect of the dis-
infectant solution may be weakened. Risk of infection to the patient
and medical personnel.
The housing plastic may become brittle and break. A color reaction
may occur with plastics.
Do not use these substances alternately.