2 Safety Instructions
80113-101_V25087
2023-11
Risk of burns for the patient under the neutral electrode!
Check in the accompanying papers of the manufacturer whether
the neutral electrode is suitable for the VIO unit used.
Use only suitable neutral electrodes.
When applying a single surface neutral electrode: Regularly check
the neutral electrode for good skin contact.
Check in the accompanying papers of the manufacturer whether
the neutral electrode cable is suitable for the neutral electrode
used.
Use only suitable neutral electrode cables.
WARNING
Positioning the neutral electrode above the heart
Risk of cardiac arrhythmia for the patient due to function-related cur-
rents from neutral electrode monitoring!
Do not position the neutral electrode over the heart or in the re-
gion of the heart.
WARNING
Incorrect application of the neutral electrode
Risk of burns to the patient!
Apply the entire contact surface of the neutral electrode to a mus-
cular part of the body with good blood circulation.
Apply the neutral electrode as close as possible to the surgical
site.
Insert the contact tab of the neutral electrode completely into the
connecting clamp. The contact tab must not touch the patient's
skin. (For reusable cord with disposable pads only.)
Align the symmetry line of the neutral electrode towards the op-
erating field. The current should flow from the active electrode
(instrument) to the symmetry line of the neutral electrode.
Check the neutral electrode regularly for good contact with the
patient's skin.
Check the neutral electrode especially when the patient has been
repositioned and after surgical steps where the unit was activated
frequently and for a long time.
Fig. 2-1