6 • Working with the electrosurgical unit
48 / 106
80113-101_V25087
2023-11
Program parameters at a glance
The following settings are specified at the factory for the programs:
Resetting to factory settings
An Erbe service technician can reset the program parameters to the factory settings
listed in the table above as required.
Important service settings
• Maximum ON time: 99 seconds
Program 1 Program 2 Program 3
(VIO
100 C only)
Program 4
(VIO
100 C only)
Socket in focus at program
start
Bipolar socket Monopolar socket Bipolar socket Monopolar socket
Footswitch assignment Bipolar socket Monopolar socket
AUTO START bipolar socket VIO 50 C: not available
VIO 100 C: not assigned
Assigned (active)
COAG mode bipolar socket BIPOLAR (= BIPOLAR SOFT COAG)
COAG power limitation bipo-
lar socket
30 watts
CUT mode monopolar socket AUTO CUT DRY CUT AUTO CUT DRY CUT
CUT power limitation monop-
olar socket
40 watts 40 watts 40 watts 40 watts
COAG mode monopolar socket SOFT COAG FORCED COAG SOFT COAG FORCED COAG
COAG power limitation
monopolar socket
30 watts 30 watts 30 watts 30 watts