EasyManua.ls Logo

Erbe VIO 50 - Page 86

Erbe VIO 50
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 Installation
86 / 106
80113-101_V25087
2023-11
If other ambient conditions must be observed for transport and
storage of the unit, you will also find them in the Technical Data.
NOTICE
Insufficient acclimatization time, unsuitable temperature during
acclimatization
If the unit was stored or transported below or above a certain tem-
perature, it will take a certain time and temperature to acclimatize.
If you do not observe the rules, the unit can sustain damage and fail.
Acclimatize the unit according to the rules in the Technical Data.
NOTICE
Overheating of the unit due to poor ventilation
If ventilation is poor, the unit can overheat, sustain damage, and fail.
Install the unit in such a way that there is an unobstructed circu-
lation of air around the housing. Installation in confined wall re-
cesses is prohibited.
WARNING
Penetration of fluid into the unit during operation or when clean-
ing the unit
Risk of electric shock to the medical personnel.
The unit may become damaged and fail.
Make sure no liquid can penetrate the unit.
Do not place vessels containing liquids on top of the unit.
If fluid has penetrated into the unit, take the unit out of operation
immediately. You can only start using the unit again once it has
been checked by a service technician.
Electrical installation
WARNING
Defective grounded power outlet, power supply network without
proper grounding, inferior-quality power cord, incorrect line volt-
age, multiple power outlets, extension cords
Risk of electric shock and other injuries to the patient and medical
personnel! Risk of damage to property.
Connect the unit to a properly installed grounded outlet.
Only connect the unit to a power supply network with proper
grounding.
Use the compatible Erbe power cord (see the Accessories, Com-
patible power cords chapter).
Check the power cord for damage. You must not use a damaged
power cord.
The supply voltage must match the voltage specified on the unit's
rating plate.
Do not use multiple power outlets.
Do not use extension cords.

Table of Contents

Related product manuals