Example 2
The flow velocity is measured.
If the flow velocity is 5 m/s, the transmitter measures in the TransitTime mode. As long as the flow velocity is > 5 m/s, the transmit-
ter has to measure in the FastFood mode.
trigger:
condition:
action:
flow velocity > 5 m/s
R1 with flow velocity > 5 m/s
measurement in the FastFood mode
Options Options
Channel A Channel A
Event trigger Remote functions
Select R1 Enable FastFood
R1
Enable R1
Yes
Source item
Flow quantities
Flow quantities
Flow velocity
Function
MAX (x>limit)
Type
Non-hold
Trigger value
5 m/s
Hysteresis
0.4 m/s
Glitch interval
1.0 s
Failure delay
1.0 s