6 Installation
FLUXUS WD 6.2 Transducers
41
UMFLUXUS_WDV1-3EN, 2018-07-16
Band clamp clasp
• Insert the tension strap into the tension strap clamp (2), see Fig. 6.25.
• Place the tension strap around the pipe and insert it into the clasp, see Fig. 6.26.
• Tighten the tension strap.
• Tighten the clasp screw.
• Tighten the nut of the tension strap clamp (2), but not too firmly in order not to damage the tension strap, see Fig. 6.25.
Quick release clasp
• Insert the tension strap into the tension strap clamp (2) and the metal spring, see Fig. 6.27 and Fig. 6.25.
• Place the tension strap around the pipe and insert it into the clasp.
• Mount the metal spring opposite to the tension strap clamp (2).
• Tighten the tension strap.
• Tighten the clasp screw.
• Tighten the nut of the tension strap clamp (2), but not too firmly in order not to damage the tension strap, see Fig. 6.25.
Fig. 6.25: Rail on the pipe
1 – tension strap clamp
2 – tension strap clamp
3–metal spring
4 – nut
5–clasp
Fig. 6.26: Tension strap with band clamp clasp Fig. 6.27: Tension strap with quick release
clasp and metal spring
1 – clasp screw 1 – clasp screw
2–metal spring