7 Connection
7.4 Service interfaces FLUXUS WD
2018-07-16, UMFLUXUS_WDV1-3EN
66
Push the cable through cap nut, compression part, basic part and sealing ring (sealing ring: only for cable gland M20, not
for cable gland 1/2 NPS).
Insert the cable into the housing of the transmitter.
Push the cable through the ferrite nut.
Prepare the cable. For the preparation of the cable, see the supplied manufacturer's documents.
Mount the connector. For the mounting of the connector, see the delivered manufacturer's documents.
Insert the connector into the Ethernet port.
Position the cable in the housing as shown in Fig. 7.14.
Fix the cable gland by screwing the cap nut onto the basic part.
Fix the transducer cable by tightening the cable gland with the ferrite nut, see Fig. 7.14.
Transmitter with aluminum housing
Remove the blind plug on the transmitter to connect the cable, see
Fig. 7.16
.
Open the cable gland of the Ethernet cable. The compression part remains in the cap nut, see Fig. 7.17.
Fig. 7.15: Cable gland
1 cap nut
2 compression part
3 basic part
4 sealing ring (only for cable gland M20, not for cable gland 1/2 NPS)
5 housing wall
6 ferrite nut
Fig. 7.16: Connection of the Ethernet cable (transmitter with aluminum housing)
1 Ethernet interface
2 foldable ferrite core
3 Ethernet connector
213456
2
1
3