9 Measurement
FLUXUS WD 9.2 Measurement settings
81
UMFLUXUS_WDV1-3EN, 2018-07-16
9.2.5.1 Output of a measured value
Select the list item Options\Outputs\...\Values.
•Press ENTER.
Output range
Select a list item.
4...20 mA
Other range
•Press ENTER.
If Other range is selected, enter the values Output
MIN and Output MAX.
The output range has to have > 10 % of the max. output
value (Output MAX). An error message will be displayed if
the output range is smaller. The next possible value will be
displayed.
Error output
An error value which is output if the source item cannot be measured, can be defined.
Select a list item for the error output, see Tab. 9.3.
Press ENTER.
If Other value is selected, enter an error value. The value has to be within the output range.
Press ENTER.
Tab. 9.2: Output of measured values, status values or event values
source item measured value event value
value status
physical quantities Flow quantities xx
Totalizers xx
Fluid properties xx
Miscellaneous
(User-def.input 1...4)
xx
Sound speed xx
Pulse xx
Diagnostic values
(except PIG detection)
x
results Diagnostic values \
PIG detection
x
Event trigger x
Source item
Values
Output range
Error value
Measuring range
Output info
Test signal
Test mea. range
see annex A, p. 156
Options\Channel A\Outputs\...\Output range
Options\Channel A\Outputs\...\Error value