9 Measurement
FLUXUS WD 9.2 Measurement settings
85
UMFLUXUS_WDV1-3EN, 2018-07-16
Output function test
The function of the output can now be tested.
Connect an external measuring instrument to the terminals of the output.
•Select Yes to test the output. Select No to display the next menu item.
Press ENTER.
Enter a test value. It has to be within the output range.
Press ENTER.
If the external measuring instrument displays the entered value, the output functions correctly.
•Select Repeat to repeat the test, Finish to display the next menu item.
Press ENTER.
•Select Yes to test the status of the output signal. Select No to display the next menu item.
Press ENTER.
Select in the scroll list Status OK or Status error.
Press ENTER.
If the external measuring instrument displays the entered value, the output functions correctly.
•Select Repeat to repeat the test, Finish to display the next menu item.
Press ENTER.
9.2.5.3 Output of an event value
Output range
Select a list item.
4...20 mA
Other range
Press ENTER.
If Other range is selected, enter the values Output MIN and Output MAX.
The output range has to have > 10 % of the max. output value (Output MAX). An error message will be displayed if the
output range is smaller. The next possible value will be displayed.
Idle state
The status of the output signal which is to be output if no event occurs is defined.
Select the value for the idle state.
Press ENTER.
Options\Channel A\Outputs\...\Test signal
Options\Channel A\Outputs\...\Enter test value
Options\Channel A\Outputs\...\Test mea. range
Options\Channel A\Outputs\...\Enter test value
Options\Channel A\Outputs\...\Output range
Options\Channel A\Outputs\...\Idle state