FT-111(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019 Page 143 of 146
- Change CR2032 battery.
- Change the main board.
Checksum error of EPL printout
format.
- Press enter key and re-load the EPL
code.
- Change E
2
ROM ( U13 ).
ADC output is not changed for the
span calibration.
- Recalibrate the scale.
- Change mainboard.
The load cell output is decreased
after loading.
- Check load cell connections.
- Check test weight loading.
- Recalibrate the scale.
- Change mainboard.
The load is not enough for span
calibration.
The load is not stable at
calibration.
- Wait until scale become stable.
- Check grounding wiring.
- Recalibrate the scale.
- Change mainboard.
Alibi memory SD card is not
installed.
- Disable Alibi memory if not required.
- Check Alibi SD card.
- Change mainboard.
Alibi SD card is not supplied from
FLINTEC.
- Install Flintec Alibi SD card.
Alibi memory serial number error.
The new alibi SD card is installed.
- Format the alibi memory SD
card. Refer to parameter 816.
Alibi memory could not be
initialized.
- Check alibi memory SD card
- Change main board.
- Check alibi memory records.
- Change Alibi memory SD card.
- Change main board.
Alibi information CSUM error.
Wrong SD card at alibi memory.
The DLC Board is broken or not
installed.
- Re-energize indicator.
- Change the DLC Board.
- Check data format of other interfaces.
- Other interfaces should not be
Modbus.
The Modbus SD card is not
installed in SD2 card slot.
- Order Modbus SD card.
- Install Flintec Modbus SD card.
Reverification the scale.
- Reverify the scale after checking the
scale hardware, load cells,
performance etc.
- Check the DLC connection (RS-485 &
Power supply) hardware.
- Check the DLC and S/N.
Shift adjustment is not available
due to load cell coefficients are out
of limits.
- Check addressing is done correctly.
- Check test weight loading on
the correct DLC.
- Check the load cell installation and
scale installation.
No regular response from load cell
- Reenergize the indicator.
- Check the DLC connection (RS-485 &
Power supply).
- Change load cell.