FT-111(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019 Page 144 of 146
E84 SN NOT MATCH
DLC yy
(1)
The address and S/N of the load
cell do not match.
- Check the DLC and S/N.
- Reenergize the indicator.
- Readdress the DLC.
E85 DLC TIMEOUT
DLC yy
(1)
- Check the DLC connection (RS-485 &
Power supply) hardware.
- Check the DLC and S/N.
E86 DLC COM ERR
DLC yy
(1)
Status error of load cell
- Check mechanical installation and
DLC.
- Check mechanical installation and
DLC.
E89 DLC CHKSUM
DLC yy
(1)
- Check termination resistors.
- Check the DLC connection (RS-485 &
Power supply) hardware.
Power supply of DLCs could not
detected.
Internal communication error
between the DLC board and the
indicator.
- Change the DLC board.
- Change the main board.
Capacity of the load cell is
different.
- Check capacity of the DLC
1) yy = Number/Address of the digital load cell.