Chapter 4: Maintenance
Autosampler Module Maintenance
CyTOF XT User Guide 101
Clean the Bottle Tray
Clean the bottle tray monthly or following any spillage.
1 Remove the bottles from the bottle tray.
IMPORTANT To prevent cross-contamination, do not allow lines from the bottle caps
to touch any fluid in the bottle tray. Let them hang in front of the system.
2 Remove the bottle tray and rinse with tap water. Dry with a Kimwipe.
3 Reinstall the bottle tray.
4 Reattach the caps and reinstall bottles.
NOTE Wipe the tubing from each cap with a separate Kimwipe moistened with DIW
before putting it back into the bottle.