Chapter 4: Maintenance
Periodic Maintenance
90 CyTOF XT User Guide
Periodic Maintenance
CAUTION Bleach, Contrad 100/Decon 90, and Citranox are caustic substances. To
prevent skin irritation or eye damage, protective gloves clothing and eye/face
protection must be worn.
NOTE Always wear non-latex powder-free gloves when working with the instrument and
handling its parts.
IMPORTANT We recommend using only Type 1 ultrapur
CyTOF XT system and its parts.
Table 6. Frequency for periodic maintenance
Component
Frequency
Carousel
Monthly or immediately after spillage
Bottle tray
Monthly or immediately after spillage
Torch assembly
Every 3 months
Air filter
Every 3 months
Nebulizer holder O-ring
Every 3 months
Nebulizer mount
Every 3 months
Autosampler air filter
Every 3 months
Table 7. Tools required for periodic maintenance and supplied by Fluidigm
Tools Provided by Fluidigm
Adjustable wrench
No. 2 Phillips screwdriver