Chapter 4: Maintenance
Clean the Injector and Spray Chamber
CyTOF XT User Guide 61
Remove the Heater Assembly
1 With the nebulizer assembly detached from the heater assembly, loosen the 2 wing nuts
by turning them counterclockwise to unlock the heater assembly from the torch assembly.
2 Gently slide the heater assembly away from the instrument.
CAUTION Finger cut hazard. Broken glass may cause injury or cutting of
fingers or hands. To avoid breakage, keep the heater assembly straight while
removing and replacing.