Chapter 4: Maintenance
Fill the Chiller with Coolant
124 CyTOF XT User Guide
Fill the Chiller with Coolant
WARNING Before performing maintenance on the chiller, switch off the RFG
power using the red circuit breaker at the right-side panel of the system. Wait at
least 5 min for residual electrical charge to dissipate.
CAUTION Hot surface hazard. The torch components, the vacuum interface, and
the heater assembly components remain hot for some time after the plasma has
been shut off. Allow sufficient time, at least 20 min, for these items to cool to room
temperature before you handle.
NOTE Wear goggles, gloves, and a lab coat.
IMPORTANT The following procedure should be done when the instrument plasma is off,
the CyTOF Software is closed, and the nebulizer is resting in the nebulizer rest after use.
1 Press and hold the button located on the top of the instrument front door for 5 sec until
a blue LED light appears and the door pops open.
IMPORTANT Before opening the instrument door, close CyTOF Software. If the
software is not closed, the alignment may shift, and the system may not reach XY
boundaries (120,000 steps). If this occurs, manually rehome the system from Instrument
Configuration > Alignment.