Chapter 4: Maintenance
Nebulizer Assembly Maintenance
108 CyTOF XT User Guide
17 Re-tighten the syringe into the valve.
18 Reinstall the plastic syringe guard and tighten screw.
19 Reconnect all color-coded the Autosampler fluidic lines as shown below.
20 Restart the Maxpar CAS Plus flow. No leaks should be observed. Continue with
acquisition.
Nebulizer Assembly Maintenance
Replace the Nebulizer Line
IMPORTANT Turn off Maxpar CAS Plus flow before removing the nebulizer line or the
nebulizer assembly.