Chapter 4: Maintenance
Autosampler Module Maintenance
CyTOF XT User Guide 97
9 Click Move To Home to raise the probe line out of the tube and back to is default
position.
Replace A Bottle Line
NOTE We recommend removing 1 line at a time to avoid mistakes.
1 Remove the nut of the damaged bottle line from the bottle.
2 Remove the bottle line from the tube guides and the hole on the Autosampler top plate.
3 Gently remove the bottle line from the groove on the syringe guard.
4 Remove the other end of the bottle line from its place on the pump and valve unit.
5 Connect the new bottle line to its position on the pump and valve unit.
6 Gently press the tube into the groove on the syringe guard.
7 Feed the bottle line through the tube guides and then through the guide hole on the
Autosampler top plate.
8 Connect the new bottle line to the bottle cap.
9 Repeat Steps 18 for additional bottle lines one at a time if necessary.