Chapter 4: Maintenance
Autosampler Module Maintenance
96 CyTOF XT User Guide
7 Use the up and down arrows until the end of the probe line is approximately 12 mm
above the bottom of the tube.
8 Click Save New Lower Position after the correct probe line position is achieved.
NOTE Use the Autosampler camera to verify the correct location of the probe line. The
images below are as examples only.