EasyManua.ls Logo

Fluidigm CyTOF XT - Recalibrate the Probe Line

Fluidigm CyTOF XT
148 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4: Maintenance
Autosampler Module Maintenance
CyTOF XT User Guide 93
2 Pull down on the quick-release connection to unlock the probe line from its position on
the stir assembly.
3 Remove the probe line from the stir assembly.
4 Thread the new probe line through the stir assembly and lock the probe line in place
using the quick release. You should hear an audible click.
5 Insert the other end of the probe line in the PROBE (gold) position of the selector valve
on the pump and valve unit and tighten by turning clockwise until finger-tight, taking
care to not leave an air gap in between the end of the line and the valve port.
6 Recalibrate the height of the new probe line as instructed below.
Recalibrate the Probe Line
1 Place an empty tube in the carousel. Note the location number.
2 In the CyTOF Software, on the Instrument Control tab, click Autosampler camera. The
tube below the probe line in the carousel is now visible.

Table of Contents

Related product manuals