Chapter 4: Maintenance
Autosampler Module Maintenance
CyTOF XT User Guide 103
a Under Actions > Push-Pump, click Down.
b Under Fill-Pump, click Down.
6 Using a No. 2 Phillips screwdriver, remove the screw from the right side of the syringe
guard on the pump and valve unit. This allows the syringe guard to be displaced from
both notches located on the left.