b Observe the spray from the nebulizer using a flashlight. It should appear as a fine
aerosol that leaves the nebulizer in an even, symmetrical V-pattern.
9 Keep the reassembled nebulizer connected and stored in the nebulizer rest for
immediate use or store the nebulizer submerged in DIW until the next use.