Chapter 4: Maintenance
Clean the Vacuum Interface Cones
CyTOF XT User Guide 73
NOTE You may need to lubricate the sampler O-rings with a small amount of
isopropanol to make the installation of the sampler cone smoother.
7 While pressing on the sampler cone with 2 fingers, detach the cone removal tool.
8 Press the edges of the sampler cone to make sure it is firmly seated taking care not to
touch the orifice.