Chapter 4: Maintenance
Reassemble the Injector and Spray Chamber
CyTOF XT User Guide 75
2 Place the spray chamber (narrow end with O-ring first) in the heater assembly. Press
down gently on the spray chamber.
3 Place the injector holder in the heater assembly cap. Ensure that THIS SIDE OUT can be
seen on the injector holder.