Remote Commands and Syntax
Commands 6
6-33
Table 6-5. Commands (cont.)
ISCE
Description
Sequential command.
Loads a byte into the Instrument Status Change Enable register described under "Check
the Calibrator Status."
Parameter The decimal equivalent of the binary number to load into the register.
Example
ISCE 56
Enables bits 3 (BOOST), 4 (RCOMP), and 5 (RLOCK) in the Service Request Enable
register.
ISCE?
Description
Sequential command.
Returns the byte from the Instrument Status Change Enable register, described under
"Check the Calibrator Status."
Parameter None
Response The decimal equivalent of the register contents byte.
Example
ISCE?
Returns: "4" if bit 3 (BOOST) is enabled (1) and the rest of the bits are disabled (0). (See
"Check the Calibrator Status" for details.)
ISCR?
Description
Sequential command.
Returns and clears the byte from the Instrument Status Change Register, described
under "Check the Calibrator Status."
Parameter None
Response The decimal equivalent of the register contents byte.
Example
ISCR?
Returns: "8" if bit 3 (BOOST) is set (1) and the rest of the bits are reset (0). (See "Check
the Calibrator Status" for details.)
ISR?
Description
Sequential command.
Returns and clears the byte from the Instrument Status Register, described under
"Check the Calibrator Status."
Parameter None
Response The decimal equivalent of the register contents byte.
Example
ISR?
Returns: "16" if bit 4 (RLOCK) is set (1) and the rest of the bits are reset (0). (See "Check
the Calibrator Status" for details.)
LCOMP_52120
Description
Overlapped command,. Ignored if not in remote. Sets inductance compensation for
attached 52120As.
Parameter
ON (enables LCOMP for all attached 52120s)
OFF (disables LCOMP for all attached 52120s)