5730A
Operators Manual
6-34
Table 6-5. Commands (cont.)
LCOMP_52120?
Description
Sequential command. Returns the current state of inductance compensation for
attached 52120As.
Parameter None
Response String
Example
LCOMP_52120?
Returns ON if LCOMP is enabled.
LIMIT
Description
Sequential command, ignored if not in remote.
Sets the maximum permissible output magnitudes, negative and positive. Both
parameters must be present, and the units must be none, volts, or amps. The units must
not conflict. If no units are specified, V is assumed.
Parameter
1. Positive limit value with optional multiplier and optional unit.
2. Negative limit value with optional multiplier and optional unit.
Example
LIMIT 1.8 A, -1.2 A
Sets the limit for ac and positive dc current to 1.8 A, and negative dc current to -1.2 A.
LIMIT?
Description
Sequential command.
Returns the programmed output magnitude limits for voltage and current.
Response
(Float, float, float, float)
1. Positive voltage limit value
2. Negative voltage limit value
3. Positive current limit value
4. Negative current limit value
Example
LIMIT?
Returns: +2.2000000E+02,-1.0000000E+02,
+1.8000000E+00,-1.2000000E+00if the programmed voltage range is -100 V to 220 V,
and the programmed current range is -1.2 A to 1.8 A.
*LRN?
Description
Sequential command.
Provides the current setting of the Calibrator. The response to this command is a string
that recreates the state of the subsequent settings when returned to the Calibrator:
Output value Whether the wideband AC Module is on Phase lock
Operate/standby Whether an auxiliary amplifier is on Phase shift
Guard Range lock Offset
Sense Two-wire compensation Scale
Output limits
Parameter None
Response
(String) A string that enables the settings that were set at the time of the *LRN?
command.
Example
SETMFC
'00000000000024400000000000408F4000000000000024400000000000002440000000
0000000000000000000030914000000000003091C00000000000807640000000000080
76C000000000000000000000000000000000000000000000000000000000000000000B
000158010000000E080000'