Remote Commands and Syntax
Commands 6
6-45
Table 6-5. Commands (cont.)
SUBNETMASK?
Description Sequential command. Returns Ethernet subnet mask for LAN communication.
Parameter None
Response String
Example
SUBNETMASK?
Returns 255.255.254.0 if the subnet mask was previously set to this value.
*TST?
Description
Sequential command, ignored if not in remote.
Initiates a series of self-tests, then returns a "0" for pass or a "1" for fail. If any faults are
detected, they are logged into the fault queue where they can be read by the FAULT?
query.
Parameter None
Response (Integer) 0 (for Pass) or 1 (for Fail)
UNCERT?
Description
Sequential command.
Returns the calculated maximum uncertainty of the Calibrator output according to the
selected calibration interval.
Parameter None
Response
1. (Float) Uncertainty of the Calibrator output (-1.0 if no specification is available)
2. (String Units for response 1 (PPM, PCT, V, A, OHM, etc.)
3. (Integer) The specification interval in days.
VOUT?
Description
Sequential command.
Returns the output amplitude expressed in volts if the currently selected units are dBm.
Parameter None
Response
1. (Float) Output amplitude
2. (String) V, A, OHM, units for response 1
3. (Float) Frequency
Example
Assuming a value of 10 dBm, 10 kHz is currently active on the main output terminals,
VOUT? returns: "2.4494897E+00,V,1.0000E+04"